User Manual

Input 
Input coupling 
DC, AC, Ground 
Input impedance 
1 MΩ ± 2% in parallel with 10 pF ± 5 pF 
Probe attenuation factor 
1X, 10X, 100X, 1000X 
Max. input voltage 
P 1280 
USB: 40 V (PK-PK) 
(DC + AC PK-PK) 
LAN: 400 V (PK-PK) 
(DC + AC PK-PK) 
P 1285 
P 1290 
400 V (PK-PK)  
(DC + AC PK-PK) 
Channel – 
channel isolation 
50 Hz: 100 : 1 
10 MHz : 40 : 1 
Time delay between 
channel (typical) 
150 ps 
Horizontal 
System 
Sampling rate range 
P 1280 
Dual CH 
0.5 Sa/s ~ 250 MSa/s 
Single CH 
0.5 Sa/s ~ 500 MSa/s 
P 1285 
Dual CH 
0.5 Sa/s ~ 500 MSa/s 
Single CH 
0.5 Sa/s ~ 1 GSa/s 
P 1290 
Dual CH 
0.5 Sa/s ~ 100 MSa/s 
Single CH 
0.5 Sa/s ~ 100 MSa/s 
Interpolation 
(sin x)/x 
Max Record length  
P 1280 
Dual CH 
≤ Max. 
sampling 
rate 
10 M 
Single CH 
P 1285 
Dual CH 
≤ Max. 
sampling 
rate 
10 M 
Single CH 
P 1290 
Dual CH 
≤ Max. 
sampling 
rate 
 5 K 
Single CH 
Scanning speed (Sa/div) 
P 1280 
5 ns/div ~ 100 s/div, 
step by 1 ~ 2 ~ 5 
P 1285 
2 ns/div ~ 100 s/div, 
step by 1 ~ 2 ~ 5 
P 1290 
5 ns/div ~ 100 s/div, 
step by 1 ~ 2 ~ 5 
Sampling rate / relay time 
accuracy 
± 100 ppm 
Interval (△ T) Accuracy 
(DC ~ 100 MHz) 
Single: 
± (1 interval time + 100 ppm x reading + 0.6 ns); 
Average > 16: 
± (1 interval time + 100 ppm x reading + 0.4 ns) 
-100- 