Set the network parameters of the oscilloscope. 
In the oscilloscope, press the Utility button and press H1 button, turn the M knob to select the LAN Set; press 
the  H2  button,  the  set  menu  is  displayed  on  the  right.  Set  the  IP  and  the  Port  to  the  same  value  as  the 
“Ports-settings” in the software in step (3). Press the H3 button, it prompts “reset to update the config”. After 
resetting  the  oscilloscope,  if  you  can  get  data  normally  in  the  oscilloscope  software,  the  connection  is 
successful.   
Fig. 62 
How to connect with the computer through a router: 
1.  Connection. Use a LAN line to connect the oscilloscope with a router, the LAN port of the oscilloscope is 
in the right side panel; the computer should be connected to the router too.   
2.  Set the network parameters of the computer. Since the oscilloscope can not support obtaining an IP 
address  automatically,  you  should  assign  a  static  IP  address.  The  Default  gateway  should  be  set 
according  to  the  router.  Here  we  set  the  IP  address  to  192.168.1.71,  Subnet  mask  is  255.255.255.0, 
Default gateway is 192.168.1.1. 