User Manual
Function Menu  Setting  Description 
operate 
Enable 
Operate 
Control enable switch 
Control operate switch 
Output 
Pass 
Fail 
Beep 
Stop 
Info 
Signal tested corresponds with the rule 
Signal tested not correspond with the rule 
Beep when it satisfies the rule   
Stop once satisfying the rule 
Control the display status of info frame     
Rule 
Source 
Horizontal 
Vertical 
Create 
Select source CH1, CH2 or Math 
Change the Horizontal tolerance value by turning the M knob 
Change the Vertical tolerance value by    turning the M knob 
Use the rule set as testing rule 
SaveRule 
Number 
Save 
Load 
Choose any one from Rule1~Rule8 as your rule name 
Click Save to save the rule 
Load some rule as the testing rule 
The description of Pass/Fail Menu is shown as the follows: 
Pass/Fail test: 
Detect whether the input signal is within the limits of the rule, if it exceeds limits of the rule, it is “Fail”; otherwise 
it is “Pass”. Also it can output fail or pass signal by built-in, configurable and photoelectric isolation output port. 
To run the test, read the following steps: 
1.  Press Utility button, then H1 button, turn the M knob to choose Pass/fail menu option, Pass/Fail menu 
will be displayed in the bottom. 
2.  Enable switch on: Press H2 button to show Operate menu, then, press F1 button to set Enable as ON. 
3.    Create rule: Press H4 button to enter Rule setting menu. Press F1 button to choose the source; Press 
F2 button, turn the M knob to set Horizontal tolerance; Press F3 button, turn the M knob to set Vertical 
tolerance; Press F4 button to create the rule. 
4.  Set output type: Press H3 button to enter Output option setting. Choose any one or two of the options 
“Pass”, “Fail” or “Beep”. “Pass” and “Fail” are mutually exclusive options,  which could not be chosen 
simultaneously. “Stop” means stop once the condition satisfies your setting. 
5.    Begin to test: Press H2 button, then F2 button to select “Start”, the test will begin. 
6.  Save rule: Press H5 button, then F2 button to save the rules, which could be called up at once when 
need, press F3 button to call up the rule saved. 
-141- 