User Manual
The description of the Save Function Menu is shown as the following table:   
Function Menu  Setting  Description   
Type 
Wave 
Setting 
Image 
Record 
Choose  the  saving  type  (the  Record  Type  see 
“27.1.  How  to  Record/Playback  Waveforms”  on 
page 134). 
When the type is Wave, the menu shows as following: 
Source 
CH1 
CH2 
Math 
Choose the waveform to be saved.   
Object & Show 
Object  1~15 
Choose the address which the waveform is saved 
to or recall from. 
Show 
ON 
OFF 
Recall or close the waveform stored in the current 
object  address.  When  the  show  is  ON,  if  the 
current object address has been used, the stored 
waveform will be shown, the address number and 
relevant  information  will  be  displayed  at  the  top 
left of the screen; if the address is empty, “None is 
saved” will be displayed. 
Save 
  Save the waveform of the source to the selected 
address. You can also press Copy button to do it. 
Storage format is BIN. 
Storage 
Internal 
External 
Save  to  internal  storage  or  USB  storage.  When 
External is selected, save the waveform according 
to the current record length (see “Record Length 
Menu”  on  page  126).  The  file  name  is  editable. 
The  waveform  file  could  be  open  by  PeakTech
®
waveform analysis software (on the supplied CD). 
When the type is Setting, the menu shows as following: 
Setting 
Setting1 
….. 
Setting8 
The setting address 
Save 
  Save  the  current  oscilloscope  setting  to  the 
internal storage 
Load    Recall the setting from the selected address 
When the type is Image, the menu shows as following: 
Save 
  Save the current display screen. The file is stored 
in BMP format and the file name is editable. The 
file can be only stored in a USB storage, so a USB 
storage must be connected first. The file is stored 
in BMP format and the file name is editable. 