3.  Repeat the steps mentioned if necessary.   
Fig. 9 Adjust Probe 
6. How to Set the Probe Attenuation Coefficient 
The probe has several attenuation coefficients, which will influence the vertical scale factor of the oscilloscope.   
To change or check the probe attenuation coefficient in the menu of oscilloscope: 
1.  Press the function menu button of the used channels (CH1 MENU or CH2 MENU). 
2.   
3.  Press H3 button to display the Probe menu; select the proper value corresponding to the probe.   
This setting will be valid all the time before it is changed again.   
 Note:  The attenuation coefficient of the probe in the menu is preset to 10X when the oscilloscope 
is delivered from the factory. Make sure that the set value of the attenuation switch in the 
probe  is  the  same  as  the  menu  selection  of  the  probe  attenuation  coefficient  in  the 
oscilloscope. 
The set values of the probe switch are 1X and 10X (see Fig. 10). 
Fig. 10 Attenuation Switch 
  Note:  When  the  attenuation  switch  is  set  to  1X,  the  probe  will  limit  the  bandwidth  of  the 
oscilloscope in 5MHz. If it is needed to use the full bandwidth of the oscilloscope, the switch 
must be set to 10X. 