43.11. How to use cursor measurement
You can measure manually the time difference between two either data in display or position difference in
memory area by cursor measurement.
Fig. 102 Cursor measurement menu
Cursor measurement time menu as below:
Function
Setting
Instruction
Increment
Time
Frequency
The time difference between two cursors.
The frequency difference between two cursors.
Cursor 1
Time
The time display of cursor 1 corresponds to trigger position.
Cursor 2
Time
The time display of cursor 2 corresponds to trigger position.
Follow up below steps to measure data pulse frequency in CH00 with cursor measurement:
1. Press “9(CURSOR) and cursor menu appears
2. Press F1till cursor display as Time”. Two purples vertical lines display in panel and refer to Cursor 1
and Cursor 2 separately.
3. Turn “CH1 POSITION” knob and set Cursor 1 position in falling edge of CH00.
4. Turn “CH2 POSITION” knob and set Cursor 2 position in next falling edge of CH00
5. Now the time increment display between two cursors is 200us and frequency is 5KHz. The time
display of Cursor 1 correspond to trigger position is 780us in “F3”. The time display of Cursor 2
correspond to trigger position is “580usin “F4”.