For example:
Measure a 16 bit data bus signal, bus will appear data of “0X99” and continuous time is 50ns. We set BUS0 to
include 16 measurement channels and sample rate as “100M”, trigger setting as duration trigger. Setting steps
as below:
1. PressTrigger MENU and menu appears.
2. PressF1” till trigger mode display as “Data Width
3. PressF2” till signal resource display as “BUS0
4. PressF3” and data in code type as red background and green digital indicator on, then insert data value
0X99”. Press “F3” again and red background disappeared. Repeat operation of step D if setting error or
need to be modified,
5. Press F4and duration in menu setting window correspond to F4 display red background. Turn CH1
Volts/Div” knob and set duration as “50ns”.
6. PressF5” till qualifier display as “>=
7. Turn Trigger level adjust knob or press “set 50%”till trigger position display as NEXT T POS = 50%”.
Then duration trigger setting finished (refer to Fig. 88) and ready for data acquisition.
Fig. 88 Data width trigger