User Manual

For example:
To set a 16 bit data Bus signal, there will have 4 value as 0X9999, 0X9998, 0X9997, 0X9996 in the bus. We
set BUS0 to include 16 measure channels and make above 4 values as trigger condition to observe the data
bus. Trigger setting follow up below steps:
1. PressTrigger MENU and menu appears.
2. Press “F1” till trigger mode display as “Sequential Queue”.
3. PressF2” till signal source display as “BUS0
4. PressF3” and data in code type as red background and green digital indicator on, then insert data value
0X9999 and set code as 0X9999”. Press F3 again and red background disappeared. Repeat
operation of step D if setting error or need to be modified,
5. Press F4 and add setting value into the trigger queue, then info window will display the value of
0X9999”.
6. Repeat the operation of step D. E, and add “0X9998”, “0X9997”, 0X9996” to trigger queue. Then
info window will display value of 0X9999”, “0X9998”, “ 0X9997”, “ 0X9996”.
7. Press “F5” and delete the queue value if queue adding in error and add again.
8. Turn “Trigger level ” adjust knob or press “set 50%”till trigger position display as “NEXT T POS = 50%”.
Then sequential queue trigger setting finished (refer to Fig. 84) and data ready for acquisition.
Fig. 84 Sequential queue trigger
-197-