43. Logic analyzer
43.1. How to set sampling system
Sampling system is to set sample rate, storage depth and filter. Different sampling setting will result in different
measure results. In the same storage depth, the higher sample rate set, the shorter the continuance time for
signal will be. If the sample rate set too low, narrower pulse signal may get lost. So the sample rate and
storage depth should be set correctly according to the actual measuring signal. Press βE (ACQU)βand display
menu as the Fig. 75:
Fig. 75 Sampling menu
Function
Setting
Instruction
Sample rate
P 1190: 20 Sa/s β 500 MSa/s
P 1230: 20 Sa/s β 1 GSa/s
Storage depth
Low memory
General
Deep memory
Storage depth of 16K
Storage depth of 256K
Storage depth of 4M
Digital Filter
None
1
2
Filter closed
Filter one pulse of sample width
Filter two pulse of sample width