User Manual

Fig. 14
17. Trigger system
Logic Analyzer is same as Oscilloscope and need to make trigger to synchronize data. The trigger system
mainly to set trigger sources, trigger mode and trigger position.
We make CH00 as trigger source and trigger mode as falling edge, trigger position in 50%. Trigger system
setting steps as below:
1. Press “Trig menu” and menu appears.
2. Press “F1” till trigger mode display as “Edge”.
3. Press “F2” or turn “CH1 Volts/Div” till trigger sources display as “CH00”.
4. Press “F3” till trigger type display as “Falling.
5. Turn “Trigger adjust” knob or press “SET 50%”till NEXT T POSwindow display as 50%”.
Then trigger system setting finished (ref to Fig. 15).
Fig. 15
-136-