6. How to Implement the Probe Compensation
When connect the probe with any input channel for the first time, make this adjustment to match the probe with
the input channel. The probe which is not compensated or presents a compensation deviation will result in the
measuring error or mistake. For adjusting the probe compensation, please carry out the following steps:
1. Set the attenuation coefficient of the probe in the menu as 10X and that of the switch in the probe as 10X,
and connect the oscilloscope probe with the CH1 channel. If a probe hook tip is used, ensure that it keeps
in close touch with the probe. Connect the probe tip with the signal connector of the probe compensator
and connect the reference wire clamp with the ground wire connector of the probe connector, and then
press the button “7(AUTOSET)”.
2. Check the displayed wave forms and regulate the probe till a correct compensation is achieved (see Fig.5
and Fig. 5).
Fig. 5 Displayed Wave Forms of the Probe Compensation
3. Repeat the steps mentioned if necessary.
Fig. 6 Adjust Probe