This oscilloscope allows you to make manual cursor measurements on time and voltage..
The cursor measurement menus are listed and described in the following table.
Function menus Settings Description
Type
OFF
Voltage
Time
Close the cursor measurement.
Display the voltage measurement cursor and menu.
Display the time measurement cursor and menu.
Delta
Display the difference of measured value between the two
cursors.
Cur1 Display the relating measured value of Cursor 1.
Cur2 Display the relating measured value of Cursor 2.
To make a voltage measurement, doing the following:
1. Press the MENU key and the function menus are displayed at the right of the screen.
2. Press the MENU or MENU key to select CURS MEAS. Four options are shown at the bottom of the
screen.
3. Press F1 key to select the measurement type Voltage. Two purple crossing dashed lines V1 and V2 are
shown on the screen.
4. Press and hold the OSC OPTION key till the / CURSOR V1 is visible on the screen. At this time,
adjust OSC or OSC and you can see that the dashed line V1 is moving up and down while the
measured voltage value of V1 relative to the zero position of the channel appears on the Curs1 item at the
bottom of the screen.
5. Press and hold the OSC OPTION key till / CURSOR V2 appears on the screen. Now, adjust the
OSC or OSC and you can observe the dashed line V2 moving and while the measured voltage
value of V2 relative to the zero position of the channel is displayed on the Curs2 item at the bottom of the
screen. Also, the absolute values of V1 and V2 can be shown on the Delta item at the bottom of the
screen.
Now, you can see a screen that looks like the following figure 45.