User Manual

9.Appendix
9.1 Appendix A: Specifications
9.1.1 Oscilloscope
Only if another instructions are provided, are all technical specifications applicable to the probe with the 10X
attenuation switch setting and the PeakTech
®
1220 digital oscilloscope. In order to be up to these
specifications, the oscilloscope should meet the following requirement.
* The instrument should operate continuously for more than 30 minutes under the specified operating
temperature.
* If the operating temperature range of variation is up to or larger than 5 Celsius degrees, the system
function menu must be opened to make the system perform a “self- calibration” procedure.
Except those specifications marked with the word Typical, all specifications can be up to.
Sampling
Sampling modes
Normal sampling
Peak detection
Average value
Sampling rate 100 MSa/s
Input
Input coupling DC, AC,Groud
Input impedance 1MΩ±2% connected in parallel with 18pF±5pF
Probe attenuation coefficient 1X, 10X, 100X, 1000X
Max. Input voltage 400V (peak)
Horizontal
Sampling rate range 0.25S/s100MS/s
Waveform interpolation sin x/x
Record length 6K points
Scanning speed rangeS/div 5ns/div100s/div, stepping in the “1-2.5-5” mode.
Time interval ( T)measurement
accuracyfull bandwidth
Single: ±(1 sampling interval time+100ppm×reading+0.6ns)
>average 16 : ±(1 sampling interval time +100ppm×reading+0.4ns)
-102