User Manual

1. Press the MENU key and the function menu appears on the right side of the screen.
3. Press MENU or MENU key to select ACQU mode, with four items selectable displayed at the
bottom of the screen.
4. Press the F3 key to select Average Factors, then, press F4 key to jump to Averaging 16 item. This
averages the outcomes of 16 acquisitions and shows the final averaging result on the screen, shown as
the following Figures.
Figure 13: Average Factor Sampling Mode
5.12. Viewing variations in a signal
You can use Persistence to observe dynamic signals.
1. Press MENU key and the function menu appears on the right side of the screen.
2. Press MENU ▲ or MENU ▼ key to select DISP SET. Four items selectable are displayed at the bottom of
the screen.
3. Press F2 key to select Persist 1 sec, 2 sec, and 5 sec, infinite or close. In this case, jump to Infinite and
the observed dynamic is kept on the screen continuously. When the item close is selected, the
Persistence function is closed.
Look at the display, a screen like the following Figure can be shown.
Figure 14: Persistence to Observe Dynamic Signals
-75-