User Manual

Function menu
Settings
Description
Format
YT
XY
Display the relative relationship between vertical voltage and
horizontal time.
Display CH1 on the horizontal axis and CH2 on the vertical axis.
Carry
Bitmap
Vectors
The data transmitted in communication are bitmaps. (Screenshots)
The data transmitted in communication are vectors. (Values)
Cymometer
ON
OFF
To set up cymometer limit to “ON” status.
To set up cymometer limit to “OFF” status.
Note: Cymometer-mode is only available at PeakTech
®
1195.
7.6.1. Display Style
The display style includes Vector and Dot displays, shown as the following Figure:
Figure 44: Dot Style
Figure 45: Vector Style
7.6.2.Persistence
With Persistence function selected, the displayed saved original data gradually decay in color and the new
data are bright in color; with infinite persistence mode selected, the oscilloscope keeps all past traces of the
displayed waveform.
7.6.3. XY Mode
This mode is only applicable to CH1 and CH2.The X-Y format plots the CH1 input as X-axis and CH2 input as
Y-axis. This display mode is convenient for viewing the phase relationship between CH1 and CH2; when the
oscilloscope is under the sampling mode in which no trigger is found, the data appear in light spots.
-98-