Connection via a router
1. Connection: Connect the oscilloscope to a router with a LAN cable. The oscilloscope's LAN port is
located on the rear panel. Now also connect the computer to the router.
2. Setting the computer's network parameters: Since the oscilloscope does not support automatic IP
address retrieval, you must assign a static IP address. The default gateway should be set according to
the router. In the following example, we set the IP address to 192.168.1.71; the subnet mask is
255.255.255.0, and the default gateway is 192.168.1.1.
3. Setting the network parameters of the PeakTech
®
oscilloscope software:
Run the software on the computer. From the Communications menu, select Port-settings. Set the "Connect
using" option to LAN. The first three bytes of the IP address are the same as the IP address in step (2). The
last byte should be different. In this example, we set the address to 192.168.1.72. The setting range of the
port number is 0 ~ 4000, but since a port lower than 2000 is always used, it is recommended to set a value
higher than 2000. In this example, we use 3000.