16. Insert C15_60
17. RW_HM Standard
18. Clamp RW_S1
19. Clamp RW_M1
20. Clamp RW_M2
21. Clamp RW_L1
22. Clamp RW_L2
23. Inner RW_TM_1
24. Clamp Dyna_S
25. Clamp Dyna_M
26. Clamp Dyna_L
27. Clamp KaYi-A
28. π-Type
M24
Select sensor installation:
0. Z method, 1. V method, 2. N method, 3. W method
M25
View distance between sensors; should be stuck to as exactly as possible
M30
Select unit system: metric or English
M31
Select flow rate unit:
Cubic Meters [m³]
Liters [l]
USA Gallons [gal]
Imperial Gallons [igl]
USA M Gallons [mgl]
Cubic Feet [cf]
USA Barrels [bal]
Imperial Barrels [ib]
Oil Barrels [ob]
Time can be per day, per hour, per minute or per second, which means 36
different unit combinations are possible
M36
Enable / disable negative totaliser
M37
1. Reset totaliser
2. Reset device to factory default settings by pressing Reset key, followed by
Back key. Be careful with this function and note down your personal
settings before using it
Password: 20140820
M39
Select the language (English or Chinese)
M40
Set damping which can be a value between 0 and 999 seconds; if “0“ is
selected, damping is disabled
M41
Set low cutoff value
M42
Set zero point; make sure no liquid passes through the pipe
M43
Delete zero point, reset to default zero point
M44
Set flow value manually (offset value); this value should normally be “0“
M45
Set scale factor; this value is set to the delivered sensors by PCE before
shipping and should only be changed after a calibration by PCE
M50
Setting the memory interval 1 … 65535 s; 0 Logger is deactivated
M52
Set Modbus RTU address 0 … 34463
M57
Number of pulses per litre
M60
View date and time (calendar for 99 years); press ENTER to make changes;
press Reset to go to the next digit
M61
View software version and serial number (ESN)
M70
Set display backlight; select how many seconds backlight will glow without
keystroke
1 … 65535 s; 0 s means display is turned off
M71
Display backlight brightness 0 … 100 %