Manual
© PCE Instruments 
25 
English 
condition that triggers the capturing when the force gauge is in curve 
capture mode. When the maximum number of data has been 
recorded or capturing was discontinued early, a capture report is 
created and saved. The curve is deleted when you leave capture 
mode. 
Baud Rate 
The baud rate for the serial interface can be set to a value between 
4800 and 230400 bps. 
This setting will only be effective after restarting the force gauge. 
Note: To make sure that all data are retrieved when the device is 
connected to a computer, the baud rate should be set as follows: 
12 Hz: ≥9600 bps 
25 Hz: ≥14400 bps 
50 Hz: ≥19200 bps 
100 Hz: ≥28800 bps 
200 Hz: ≥38400 bps 
400 Hz: ≥57600 bps 
800 Hz: ≥115200 bps 
1600 Hz: ≥230400 bps 
Due to the limited speed of serial interfaces, some data get lost when 
transferred to a PC if the sampling rate is higher than 800 Hz. The 
readings will, however, not get lost in the device. 
Display Angle 
Here, you can set the display angle. You can select 0 or 180 °. 
Auto Backlight 
This function also reduces the energy consumption. The brightness of 
the display backlight will automatically be turned off when the meter 
has not been used for a certain period of time. 
Now Voltage 
This window shows the current battery level. 
Reset 
You can reset the device to factory default settings, e. g. if you have 
made an incorrect setting or if other problems with the settings 
occurred. 
Language 
In this window, you can change the output language. English or 
German can be selected. 
Connection 
In this window, you can see the circuit diagram of the input port (see 
chapter 3.2). 