Manual
© PCE Instruments 
24 
English
Function 
Description page 2 
Capture Length 
You can set a value between 1 and 1280 seconds. This value 
represents the duration of curve capturing in capture mode which 
depends on the sampling rate: 
Sampling rate 6 Hz: 1 ~ 1280 seconds 
Sampling rate 12 Hz: 1 ~ 640 seconds 
Sampling rate 25 Hz: 1 ~ 320 seconds 
Sampling rate 50 Hz: 1 ~ 160 seconds 
Sampling rate 100 Hz: 1 ~ 80 seconds 
Sampling rate 200 Hz: 1 ~ 40 seconds 
Sampling rate 400 Hz: 1 ~ 20 seconds 
Sampling rate 800 Hz: 1 ~ 10 seconds 
Sampling rate 1600 Hz: 1 ~ 5 seconds 
F/P Boundary
You can set a value between 1 and 99999. This setting is used during 
peak value measurement to determine the first peak value. When you 
press the Arrow Right key, a new peak value measurement will start. 
Meanwhile, the values peak-to-peak (Vmax), valley-to-peak (Vmin) 
and new peak (Vnew) are updated continuously. For example, if 10 is 
set as the criterion, Vmax or Vmin will be counted as the first peak 
value when the absolute value of (Vmax - Vnew) or (Vmin - Vnew) is 
above 10. 
Serial Port 
This port is used to control the real-time data transfer. The following 
parameters can be set: 
Prohibit: The real-time data transfer of the serial interface is 
prohibited. 
Key/Order: A single output will take place when you press the Up key 
or when an output command is received. When the force gauge is 
connected to a computer, the programmes on the computer will 
automatically disable the output function. 
Change: A single output will take place when the measuring data 
change. 
Stabilize: A single output will take place when the reading stabilises. 
Consecutive: The measuring data are transferred without interruption. 
Auto Power Off 
This function reduces the energy consumption. The force gauge will 
automatically power off when it has not been used for a certain period 
of time. 
Max Charge V 
This window shows the maximum voltage of the battery. 
Clear Storage 
Here, you can delete saved measurement reports and curves.  
Important information: 
When the memory is full, all data will automatically be deleted to 
enable new data to be saved. 
Factory Test 
Only relevant for customer service. 
S/N 
This window shows the serial number of the device which cannot be 
changed. 
Capture Trigger 
Here, you can set a value between -99999 and +99999. The range of 
values depends on the set unit. This parameter stipulates the 