4.2  Settings 
When you are in measurement mode, press the OK key to enter the settings screen which is divided 
into 2 pages: 
Page 1          Page 2 
In order to change settings, select the menu item with the arrow keys and confirm with the OK key. 
The values can then be changed by  means  of the arrow keys. Then  press “OK” to  confirm the 
settings or the Back key to discard. 
Function 
Description page 1 
Display Unit 
The display unit can be set to: „N“, „kg“, „lb“ or „KPa“ 
Force Area 
The force area can be set to a value between 999.99cm
2 
and 0.01cm
2
and is included in the calculation if the display unit selected is „kPA“ 
(important for the accuracy). 
Zero Tracking 
For zero tracking, you can set: 
„Off“, „0.1 N“, „0.2 N“, „0.3 N“, „0.4 N“, „0.5 N“ 
Values below the value set are automatically excluded before the zero 
point stabilises. After stabilisation of the reading, the sampling rate will 
be 1 x per second. Deviations from the measured value which are 
below the set value are automatically excluded in order to keep the 
displayed values. 
Sampling Speed 
You can set how many measurements are taken per second. A value 
between 6 and 1600 Hz can be set here. 
Note: The higher the sampling rate, the lower the accuracy will be. 
Higher sampling rates are suitable for dynamic measurements whereas 
lower sampling rates are more suitable for static and slow 
measurements. 
Calibrate Grav 
Enter the gravity at the place of calibration. 
Alarm Upper 
The upper alarm can be set to +/- 9999.9. 
Alarm Mode 
You can choose “Within“ (within alarm limit), “Beyond“ (outside alarm 
limit), “Fracture“ (overload alarm) or “Off“. 