7  Calibration 
If the device has not  been used for a longer period of  time, it must  be calibrated carefully. A 
calibration is also necessary if the impact devices are changed frequently. Use the hardness test 
block to calibrate the meter. 
-  To enter calibration mode, switch on the instrument by pressing the On/Off key  . 
Then  press  and  hold  the  arrow  key   until  the  device  indicates  that  it  is  in 
calibration mode. 
-  Measure five different points on the standard test block. You will see in the display how 
many  measurements  you  have  made.  After  the  fifth  measurement,  the  display  will 
directly show the average value of the calibration. 
-  Compare the average value with the HLD value on the upper side of the standard test 
block. 
-  Use the arrow keys   and   to change the average value of the calibration. 
The calibration range is ±150HL. 
-  Press the Menu key   to confirm the calibration or the Back key   to cancel 
the calibration. 
Note: Measuring parameters such as material, hardness scale and impact direction cannot be 
changed during the calibration. 
8  Maintenance 
8.1  Storage 
Store the measuring instrument in its carrying case and make sure the environmental conditions 
are within normal ranges. Avoid magnetic fields, corrosion and shocks. 
8.2  Removing the sensor cable from the meter 
Hold  the  moveable  connecting  piece  on  the  sensor  cable  which  is  inserted  into  the  sensor 
connection and pull it away from the instrument. 
8.3  Maintenance and care of the impact device 
-  After using the impact device approx. 1000 – 2000 times, the guide tube and the impact 
body should be cleaned by using the nylon brush that comes with the meter. To do so, 
unscrew the support ring, then take out the impact body, move the nylon brush through 
the guide tube counterclockwise, making helical movements until you reach the bottom. 
Repeat this procedure four times. Re-install the impact body and the support ring. 
-  Do not use any lubricants inside the impact device. 
-  Place the impact body back into the loading tube and screw the support ring back onto 
the end of the impact device. 