6.2  Measurement 
- Hold the impact device between 
 your thumb and your index finger 
 and place it on the sample. Make 
 sure that it is placed firmly on the 
 surface and that the impact 
 direction is vertical to the test 
 surface. Otherwise, incorrect 
 measurement values can occur. 
- Use one hand to press the 
 support ring of the impact device 
 onto the sample. Make sure that 
 the support ring lies firmly on the 
 surface and that the impact 
 direction is vertical to the test 
 surface. Use your other hand to 
 hold the handle of the impact 
 device between your thumb and 
 index finger. Push the handle 
 towards the support ring up to 
 the stop. Then return the handle 
 to its original position but do not 
 stop holding it while doing this. 
- During this movement, the 
 impact body in the tube of the 
 sensor will be placed in its 
 original position. 
-  By pressing the trigger button on top of the moveable handle, the measurement is 
started. The impact body in the sensor tube strikes the test surface. 
-  The reading is displayed on the LC display. 
-  The distance between any two impact points or between the centre of any impact point 
and the edge of a sample must comply with the requirements in the following chart: 
Impact 
device 
Centre-to-centre distance 
of two impact points 
Distance from centre of impact point to 
edge of sample 
D 
≥3 mm 
≥5 mm 
-  On any surface to be measured, you should test at least five different measuring points 
to calculate the average. 
6.3  Evaluation of the measurements 
After each measurement, the measured value will be indicated in the display. 
The PCE-2000N calculates and saves an average value. Therefore, various measuring positions 
should be tested. As soon as the pre-set number of measurements is reached, the tester will 
make a sound and the display will show the average value. 