User's Manual

6
Photometer 012-05631A
Procedure
NOTE: You may want to cover the crossed-arrow objects on each light source to reduce
the excess light in the room. The room lights must be off for this experiment.
Set the neutral density filter for 100% transmission.
While looking into the photometer’s conical eyepiece, adjust the position(s) of the light
source(s) until the two sides of the orange indicator have equal intensity.
Set the neutral density filter for 75% transmission.
While looking into the photometer’s conical eyepiece, rotate the label-side polarizer until the
two sides once again have equal intensity. Record the angle in Table 2.1. Rotate the polarizer
back to zero and repeat the measurement two more times.
Repeat the previous step for 50% and 25% transmission.
Table 2.1: Data and Results
% transmittance 75% 50% 25%
trial 1
trial 2
trial 3
Average angle φ
cos
2
φ
% difference
Analysis
For each of the neutral density filter settings, calculate the average of the three trials and
record the average angle in Table 2.1.
To calculate the predicted percentage transmittance for each case, calculate the square of the
cosine of each average angle and record in Table 2.1.
Calculate the percent difference between the percentage transmittance and the predicted value
for each case and record in Table 2.1.