User's Manual
SE-9639 Franck-Hertz Experiment
4
012-14264A
Replace the Argon Tube
• Use a flat-blade screwdriver to remove the two small screws that hold the back
plate onto the argon tube enclosure.
• Use a small flat-blade screwdriver to pry the back panel off of the enclosure.
• Pull up on the elastic pressing spring and rotate it off the argon tube.
• Gently pull out the argon tube.
• Then, install a new tube and replace the elastic pressing spring.
• Finally, close the case and replace the two small screws.
Fuse Replacement
• Disconnect the power cord from the instrument.
• Open the fuse cover and remove the fuse. (The fuse is inside a tray. Use a small
screwdriver or other tool to pry the tray open.)
• Replace the fuse(s). Use the same type of fuse (250 V T2A).
• Reconnect the power cord and turn on the instrument.
• If the problem persists, contact Brolight Corporation for service.
Argon Tube Specifications
Filling gas argon
Filament voltage 6.3 V DC
Accelerating voltage 100 V DC
Wave crest (or trough) number 6
Life span 2000 hours
• Note: The tube is a thin-walled, evacuated glass bulb. Handle with
care! Do not expose the tube to mechanical stress or strain.
Note: Replace the argon tube with the same type: Model SE-9645 Franck-Hertz Ar-Tube.
The fuse is inside a
tray. Open the
cover to remove
the fuse.
WARNING
To reduce the risk of electric shock or
damage to the instrument, turn the
power switch OFF and disconnect the
power cord before replacing a fuse.
Fuse Cover Tray
Pry
here
Note: Replace the burned fuses with new fuses of the same type. (One spare fuse is included.)