User's Manual

Model No.PS-2116A Appendix: Calibration
5
012-09688B
In the Conductivity Sensor screen, touch “Conductivity
(10X Probe) to select that choice, and then touch
“Show”.
A graph display of Conductivity and Time opens. Touch
the “Experiment Tools” icon in the lower right corner.
The “Experiment Tools” screen opens. Touch “Calibrate
Sensor” to select that choice and then touch “OK”.
The “Calibrate Sensor: Select Measurement” screen
opens. and the default selections are “Conductivity (10X
Probe) (S/cm)” as the “Measurement” and “1 point
(Adjust Slope Only)” as the “Calibration Type”. Touch
“NEXT”.
The “Calibrate Sensor: Enter Values” screen shows the
“Calibration Point 2” menu.
Touch “Standard Value:” to open the keypad. Use the
keypad to enter the standard value for the conductivity
of the standard conductivity solution.
When the number is entered, touch “OK” on the keypad
to return to the “Calibrate Sensor” screen. Touch “OK”
to finish the calibration.
First, touch “Conductivity
(10X Probe)” to select it:
Then, touch “Show”:
Tou c h
“Experiment Tools”:
First, touch
“Calibrate Sensor”
to select it:
Then, touch “OK”: