User's Manual
®
Ocean Optics with Xplorer GLX Spectrometer Analysis Configuration Screen
4
Spectrum Preview
When the configuration screen opens, the GLX automatically starts
monitoring and displaying spectrum data in the upper half of the screen.
With this preview, you can immediately see the effect of changes that
you make to parameters. Press to manually stop or re-start spec-
trum monitoring.
Function Keys
Use the function keys to save and clear dark and reference scans and to
close the configuration screen.
F1 Save Dark With the light source off or blocked (so that no light
enters the spectrometer), press to save the dark scan.
F2 Save Ref With “reference light” entering the spectrometer,
press to save the reference scan. Do this after setting the parame-
ters in the Data Acquisition tab (see “Data Acquisition Tab” below).
Your “reference light” may be one of the follow (depending on what
type of measurement you plan to make):
• The Integrated Light Source with a reference cuvette
2
in place (for
measuring absorbance and transmission of light through a solution).
• A lamp with a known color temperature (for measuring the relative
irradiance of light sources).
• Any light source (if you plan to use only the Intensity and Minus
Dark measurements).
F3 Clear Scan Press to delete the dark and reference scans.
F4 Close Press to close the configuration screen.
Navigating in the Configuration Screen
The configuration parameters and options are contained in three tabs:
Data Acquisition, Lamp, and Time Acquisition.
To navigate between the tabs, press the left and right arrow keys. (Note
that the Time Acquisition tab will not be available until you have saved
dark and reference scans.)
Navigate within a tab by pressing the up and down arrow keys to high-
light different parameter and options. To change the highlighted param-
eter or option, do one of the following:
• Press or to change the value of the parameter by 1.
• Press to make the parameter editable. Type the desired
value. Press again.
• Press to toggle a two-choice option.
or
• Press to open a pop-up menu. Use the arrow keys to highlight
the desired menu option. Press again.
Spectrum
preview
F4
F3F2
F1
Function keys
F1
F2
2
A reference cuvette represents 100%
transmission.
F3
F4
Press and to move between tabs.
Press and to
move within a tab.
+