User's Manual

®
Ocean Optics with Xplorer GLX Spectrometer Analysis Configuration Screen
3
E. Configure the Spectrometer
In the Spectrometer Analysis Configuration screen, follow these general
steps. (For a detailed explanation of the parameters in the configuration
screen and how to change them, see pages 3–6.)
1. Set the parameters in the Data Acquisition tab as desired.
2. (This step is not always necessary.) Go to the Lamp tab and change
the parameters there if necessary.
3. (Do this step if you plan to use Time Acquisition or Beer’s Law
mode or the Minus Dark, Absorbance, Transmittance, or Relative
Irradiance measurements.)
a. Turn off the light source or block light from entering the spec-
trometer. Press to save the dark scan.
b. Allow “reference light” to enter the spectrometer. Press
to save a reference scan.
1
4. (Do this step only if you plan to use Time Acquisition or Beers Law
mode.) Go to the Time Acquisition tab. Set the scan mode and other
parameters as desired. (See page 6 for more information.)
5. After you have finished setting spectrometer parameters and
options, press to close the configuration screen.
See page 6 for instructions on recording and viewing data.
Spectrometer Analysis Configuration Screen
The Spectrometer Analysis Configuration screen is where you set the
parameters that control how the spectrometer and GLX collect and pro-
cess data.
To Open the Configuration Screen
If the spectrometer is not yet connected:
Connect the spectrometer to the GLX's USB port and wait for the
configuration screen to open automatically.
If the spectrometer is already connected:
1. Press + to open the Graph screen (if it is not already
open).
2. Press to open the Tools menu.
3. Use the arrow keys to highlight Spectrum Analysis Config.
4. Press .
Spectrometer Analysis
Configuration screen
(appears after the Initializing screen)
F4
F3F2
F1
F1
F2
1
What you use as the “reference light
depends on what you plan to measure.
See see “F2 Save Ref” on page 4 for
details.
F4
F1
F3