Meter Installation in 4838 Reactor Controller
1. Unplug the power cord of the 4838 controller. Remove the (2) screws located on
the top/front cover at the rear corners of the controller. Gently lift the cover
forward, which is hinged at the bottom, taking care not to apply tension on any
internal wiring.
2. If applicable, remove and discard the
black hole plug on the sloped front
panel. This can be done by pinching the
clasps on the back of the plug from the
inside of the front plate.
3. Insert the meter (2082E) through the
panel cutout, from outside in, making
sure the rubber gasket is on the outside of the controller front panel. (If wires
are preinstalled, remove the terminal connector by releasing the tabs on the
sides of the module and gently pulling the terminal connector off. Reference
photos below.) Slide the white mounting bracket onto the module with the
mounting screws pointing towards the front panel. Slide the mounting bracket
forward until it touches the controller panel. The clasps on the mounting bracket
should align with the groves on the module. Gently tighten the screws until the
module is held into place. (If removed, reattach the terminal connector.)
Terminal Connector Removal
Installation Instructions for
External Temperature Limit Module (ETLM)