User Manual
4848B Reactor Controller
Parr Instrument Company
8
A load fuse is mounted on the back panel on 115V 
systems. This fast-acting, 250VAC, 20 amp fuse is in-
tended to protect the controller and prevent electric 
shock hazard from fault from occurring.
Two load fuses are mounted on the back panel 
on 230V systems. These are fast-acting, 20 Amps, 
250VAC and are intended to protect the controller 
and to prevent electric shock hazard from fault from 
occurring.
Caution!
Unplug unit before servicing. For continued 
protection against possible hazard, replace 
fuses with same type and rating of fuse.
Back Side
4848B Reactor Controller
Protective Fuses and Relays
An instrument fuse is mounted on the back panel. 
This is a Fast-acting, 250VAC, 1.0 amp fuse, intended 
to protect the temperature control module and any 
installed display modules.
The motor reset/motor fuse provides overload 
protection for 0-90VDC motors. This is a re-settable 
breaker designed to protect the motor if it becomes 
overloaded. The breaker will have a rating selected 
to match the motor on the apparatus.
A lighted fuse holder provides protection for 0-180 
VDC motors. A light will come on if the motor be-
comes overloaded and the fuse burns out. The fuse 
rating will be selected to match the motor on the 
apparatus.