Operator's Manual

Table Of Contents
22
08424705001 (04) 2021-06 USA • Accu-Chek Inform II Operator’s Manual Version 7.2
Shut down meter In standby mode (“Automatic power off”), energy contin-
ues to be drawn from the battery and it depletes within a
day if not docked. Shutting down the meter shuts down
wireless communication and all other functionalities.
Date and time, however, are maintained. Powering up the
meter will take slightly longer than from power-off
(standby) mode.
To shut down the meter, press the On/Off button for
about 5 seconds and release the button as soon as the
Roche logo is displayed and the meter beeps. The screen
goes blank and the meter is shut down.
Use meter shutdown when you want to remove or replace
the battery pack (see page 122).
If you press the On/Off button for too long, a meter reset
will be triggered after about 12 seconds (see page 181)
and date and time will be lost.