Operator's Manual

Table Of Contents
Appendix • A
201
08424705001 (04) 2021-06 USA • Accu-Chek Inform II Operator’s Manual Version 7.2
Result deletion
Result deletion algorithm. Automatic or
First in First out.
0: automatic
1: FIFO
1NY
Result retention time: Number of days
since a result was measured before it will
be deleted automatically if result deletion
algorithm is set to “automatic”. Results are
then deleted when powering up the meter
or performing a test.
1 – 1,000 d 30 N Y
Result download required 0: disabled
1: enabled
0NY
Electronic communication
Download warning 0 – 999 h 0: disabled N Y
Download Lockout 0 – 999 h 0: disabled N Y
Maximum number of list items transferred
in one POCT1-A message
1 – 500 75 N Y
Application timeout (within this time the
application expects a response by the DMS
to any POCT1-A command)
30 – 3,600 s 60 N Y
TLS Encryption of communication 0: disabled
1: enabled
0NY
Compression of data communication 0: disabled
1: enabled
0NY
Subject/Attribute Range Default Device DMS