Appendix • A
199
08424705001 (04) 2021-06 USA • Accu-Chek Inform II Operator’s Manual Version 7.2
Patient ID
Patient ID confirmation 0: disabled
1: name
2: date of birth (DOB)*
3: name/DOB*
0NY
* Depends on available DMS settings.
Patient ID entry mode 0: Keyboard / Scan
1: List / Keyboard / Scan
2: Scan only
3: Prompt (numeric)
0YY
Patient ID validation
(Any character allowed [except non-print-
able ones, represented as hexadecimal
ASCII values by 0x01 - 0x1F and 0x7F]. Via
the meter keyboard only A-Z/a-z, 0-9, “.”
[period], “-” [hyphen] can be entered.)
Remark: leading and trailing spaces are
truncated.
0: none
1: length
2: list
3: list allowing entry if not
on list
4: length if numeric
0YY
Patient ID maximum length (used to
validate a patient ID if patient ID validation
mode is set to “Length” or “List - allowed
to add new”)
0-20 20 Y Y
Patient ID minimum length (used to vali-
date a patient ID if patient ID validation
mode is set to “Length” or “List - allowed
to add new”)
0-20 0 Y Y
Patient name length 0-25 characters N/A N N
Subject/Attribute Range Default Device DMS