Operator's Manual
Table Of Contents
- What is new in publication version 7.2?
- 1 Introduction
- 1.1 Before you start
- 1.2 Important safety instructions and additional information
- Important information regarding safety
- Disposal of the system
- Product safety
- General care
- Accessory box
- Meter
- Power off meter
- Automatic power-off
- Shut down meter
- Automatic shutdown
- Battery Pack
- Touchscreen
- Electromagnetic compatibility (EMC)
- Electrostatic discharge (ESD)
- Wireless connectivity
- Radiofrequency radiation exposure information
- Local Area Network: protection from unauthorized access
- Wired network connection
- 1.3 System components
- 1.4 Overview of the meter
- 1.5 Overview of the code key reader
- 1.6 Overview of the Accu-Chek Inform II Base Unit
- 1.7 Overview of the Accu-Chek Inform II Base Unit Hub
- 1.8 Overview of the accessory box
- 1.9 Reagents and consumables
- 1.10 Instructions for initial setup
- 2 Powering Up and Entering an Operator ID
- 3 Patient Glucose Testing
- 3.1 Information regarding blood glucose testing
- 3.2 Performing a patient glucose test
- Overview of test procedure
- Entering or selecting the patient ID
- Entering the patient ID manually
- Selecting the patient ID from a list
- Entering a patient ID with barcode scanner
- Confirming or selecting the test strip lot
- Patient identification information
- Inserting test strips
- Obtaining a blood sample
- Applying a blood sample
- Results screen
- Adding comments
- Add barcode content to a result
- Additional Patient Test
- 4 Glucose Control Testing
- 5 Review Results
- 6 Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
- 7 Linearity Testing
- 8 Proficiency Testing
- 9 Initial Startup
- 9.1 Connecting the base unit
- 9.2 Installing or replacing the battery pack
- 9.3 Docking the meter
- 9.4 Initial setup on the meter
- 9.5 Opening the Setup Menu
- 9.6 Date and time format
- 9.7 Display options and optional tests
- 9.8 Options for test strips
- 9.9 Options for glucose control tests
- 9.10 Value ranges (normal, critical, reportable)
- 9.11 Options for Operator ID Entry
- 9.12 Patient ID options
- 9.13 Creating a setup password
- 9.14 Setting the date and time
- 9.15 Beeper options
- 10 Maintenance and Care
- 11 Troubleshooting
- 12 General Product Information
- A Appendix
- B Appendix
- C Supplement for Other Test Entry
- D Supplement for Observed Test Sequence
- E Supplement for Enhanced Workflows
- F Appendix for Accu-Chek Inform II Base Unit (legacy version)
- G Limited Warranty
- Index
Appendix • A
193
08424705001 (04) 2021-06 USA • Accu-Chek Inform II Operator’s Manual Version 7.2
A Appendix
A.1 Table of configuration options
This section provides an overview of all the settings
available.
The two rightmost columns describe the accessibility of a
configuration parameter on the device (Setup Menu) and
via the data management system (DMS).
“Y” (Yes) means, that this parameter is available, “N” (No)
means, that this parameter is not available using the
respective configuration method.
Configuration options
DEVICE: If the meter is configured by a system admin-
istrator using a data management system, the configu-
ration options on the meter may be disabled to avoid
conflicts in the settings. This option is indicated by the
use of parentheses e.g., (Y).
DMS: Configuration options may vary depending on
available DMS settings.
Subject/Attribute Range Default Device DMS
Timer
Power-off timeout (meter powers off
automatically if no activity)*
30 – 3,600 s 300 N Y
Shut down (meter automatically shuts down
30 minutes after powering off)
0: disabled
1: enabled
1NY
Time Interval reconnect (0 = disabled) 0: disabled
10 – 1440 minutes
(24 hrs)
30 N Y
* Not applicable in measurement mode (see “Automatic power-off” on page 21).