Operator's Manual

Table Of Contents
Troubleshooting • 11
177
08424705001 (04) 2021-06 USA • Accu-Chek Inform II Operator’s Manual Version 7.2
E-FCG Invalid Scan: Invalid Scan. The scan returned an entry > 20 characters. Please see your System
Administrator.
E-FCH Invalid Range: The range entered must be between 10 and 600 mg/dl. Please re-enter.
E-FCK Invalid Range: The Normal Range Hi value must be greater than the Normal Range Lo value
(currently *). Please re-enter.
<* is the Normal Range value defined by the institution on the device>
E-FCL Invalid Range: The Critical Range Hi value must be greater than the Critical Range Lo value
(currently *). Please re-enter.
<* is the Critical Range value defined by the institution on the device>
E-FCM Invalid Range: The Reportable Range Hi value must be greater than the Reportable Range Lo
value (currently *). Please re-enter.
<* is the Reportable Range value defined by the institution on the device>
E-FCN Invalid Range: The Reportable Range Lo value must be less than or equal to the Critical Range Lo
value (currently *). Please re-enter.
<* is the Critical Range value defined by the institution on the device>
E-FDB Invalid Range: The Reportable Range Hi value must be greater than or equal to the Critical Range
Hi value (currently *). Please re-enter.
<* is the Critical Range value defined by the institution on the device>
E-FDC Invalid Range: The Critical Range Hi value must be greater than or equal to the Normal Range Hi
value (currently *). Please re-enter.
<* is the Normal Range value defined by the institution on the device>
E-FDD Invalid QC Hours: QC Hours must be greater than 0 and less than 10,000. Please re-enter.
E-FDF Invalid Strip Limit: QC Strips must be greater than 0 and less than 1,000. Please re-enter.
E-FDH Invalid Strip Lot Barcode: The scanned in barcode is invalid for a Strip lot barcode. Please scan in
a valid barcode, or see your System Administrator
E-FDK Undefined Strip Lot: The barcode scanned strip lot: * is not defined on this meter. Please see your
System Administrator.
<* is the test strip lot number>
E-FDL Invalid Limit: The length entered must be between 0 and 20. Please enter a valid length.
E-FDM Invalid Limit: The Operator ID Max length must be greater than or equal to the Operator ID Min
length (currently *). Please re-enter.
<* is the Operator ID minimum length defined by the institution >
E-FDN Invalid Limit: The Patient ID Max length must be greater than or equal to the Patient ID Min length
(currently *). Please re-enter.
<* is the Patient ID minimum length defined by the institution >
E-FFB Invalid Operator Barcode: The barcode scanned is not a valid operator barcode. Please scan a
valid operator barcode or see your System Administrator.
Error Messages
Code Text