Operator's Manual
Table Of Contents
- What is new in publication version 7.2?
- 1 Introduction
- 1.1 Before you start
- 1.2 Important safety instructions and additional information
- Important information regarding safety
- Disposal of the system
- Product safety
- General care
- Accessory box
- Meter
- Power off meter
- Automatic power-off
- Shut down meter
- Automatic shutdown
- Battery Pack
- Touchscreen
- Electromagnetic compatibility (EMC)
- Electrostatic discharge (ESD)
- Wireless connectivity
- Radiofrequency radiation exposure information
- Local Area Network: protection from unauthorized access
- Wired network connection
- 1.3 System components
- 1.4 Overview of the meter
- 1.5 Overview of the code key reader
- 1.6 Overview of the Accu-Chek Inform II Base Unit
- 1.7 Overview of the Accu-Chek Inform II Base Unit Hub
- 1.8 Overview of the accessory box
- 1.9 Reagents and consumables
- 1.10 Instructions for initial setup
- 2 Powering Up and Entering an Operator ID
- 3 Patient Glucose Testing
- 3.1 Information regarding blood glucose testing
- 3.2 Performing a patient glucose test
- Overview of test procedure
- Entering or selecting the patient ID
- Entering the patient ID manually
- Selecting the patient ID from a list
- Entering a patient ID with barcode scanner
- Confirming or selecting the test strip lot
- Patient identification information
- Inserting test strips
- Obtaining a blood sample
- Applying a blood sample
- Results screen
- Adding comments
- Add barcode content to a result
- Additional Patient Test
- 4 Glucose Control Testing
- 5 Review Results
- 6 Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
- 7 Linearity Testing
- 8 Proficiency Testing
- 9 Initial Startup
- 9.1 Connecting the base unit
- 9.2 Installing or replacing the battery pack
- 9.3 Docking the meter
- 9.4 Initial setup on the meter
- 9.5 Opening the Setup Menu
- 9.6 Date and time format
- 9.7 Display options and optional tests
- 9.8 Options for test strips
- 9.9 Options for glucose control tests
- 9.10 Value ranges (normal, critical, reportable)
- 9.11 Options for Operator ID Entry
- 9.12 Patient ID options
- 9.13 Creating a setup password
- 9.14 Setting the date and time
- 9.15 Beeper options
- 10 Maintenance and Care
- 11 Troubleshooting
- 12 General Product Information
- A Appendix
- B Appendix
- C Supplement for Other Test Entry
- D Supplement for Observed Test Sequence
- E Supplement for Enhanced Workflows
- F Appendix for Accu-Chek Inform II Base Unit (legacy version)
- G Limited Warranty
- Index
176
08424705001 (04) 2021-06 USA • Accu-Chek Inform II Operator’s Manual Version 7.2
E-CBK Invalid Password: Invalid Setup password. Please enter valid password or see your System
Administrator.
E-CBL No Operators Loaded: There are no operators loaded in this meter. In order to continue, operators
must be downloaded. Please see your System Administrator.
E-CBM QC not allowed: The Operator ID entered was not found in the list of valid operator IDs for QC.
Please see your System Administrator.
E-CBN Invalid Observer ID: Observer ID is invalid. ID entered was not found in the list of valid observer
IDs. Please see your System Administrator.
E-CCB Invalid Observer ID: Invalid Observer ID: The observer ID is invalid. The observer supervising the
current OTS test must log in again.
E-DNL Invalid Time: Time entered is in the future. Please enter the time in HH:MM AM/PM format. (If in
24 hr mode the AM/PM is removed.)
E-DNM Invalid Date: Date entered is beyond the code key expiration date. Please re-enter date.
E-DNN Invalid Date: Date entered is in the future. Please enter the date in MM/DD/YY format.
E-FBB Invalid Date: Date entered is expired. Please check the package and re-enter date.
<for control and linearity lots expiry dates>
E-FBC Invalid Time: Time entered is invalid. Please enter the time in HH:MM AM/PM format.
E-FBD Invalid Date: Date entered is invalid. Please enter the date in MM/DD/YY format.
E-FBF Invalid Patient ID: Patient ID is invalid. ID entered was not found in the list of valid patient IDs.
Please see your System Administrator.
E-FBG Invalid Patient ID: Patient ID length is invalid. ID must be between * characters long.
<* is the minimum and maximum character length defined at the institution >
E-FBH Invalid Patient ID: Patient ID length is invalid. ID must be * characters long.
<* is the minimum character length defined at the institution >
E-FBL Comment required: A comment is required for this result. Please enter at least one comment.
E-FBM Invalid Lot Number: A valid lot number must be provided. Please enter a lot number.
E-FCB Invalid Control Range: Level 1 (Lo) Max must be greater than Level 1 (Lo) Min (currently *). Please
re-enter.
<* is the L1 Control Range minimum value defined by the Control Lot information on the device>
E-FCC Invalid Control Range: Level 2 (Hi) Max must be greater than Level 2 (Hi) Min (currently *). Please
re-enter.
<* is the L2 Control Range minimum value defined by the Control Lot information on the device>
E-FCD Invalid Control Range: Level 2 (Hi) Min must be greater than Level 1 (Lo) Max (currently *). Please
re-enter.
<* is the L1 Control Range maximum value defined by the Control Lot information on the device>
E-FCF Invalid Sample ID: Invalid sample ID, length is invalid. ID must be at least one character. Please
re-enter.
Error Messages
Code Text