Operator's Manual

Table Of Contents
Troubleshooting • 11
173
08424705001 (04) 2021-06 USA • Accu-Chek Inform II Operator’s Manual Version 7.2
Warning; touch to confirm.
Error; touch to confirm. To resolve the problem, take
the actions suggested.
Sample message type Description
Loss of Date/Time
[W-SBD]
D/T settings lost due to
meter reset or loss of
power
. Before performing
further tests: Enter D/T
manually and synchronize
meter. OR: Ask System
Admin. for help.
123
Operator ID
12:48 pm
03/17/20
Invalid Scan
Invalid Scan. The scan
returned an entry > 20
characters. Please see
your System Administrator
.
[E-FCG]