Operator's Manual
Table Of Contents
- What is new in publication version 7.2?
- 1 Introduction
- 1.1 Before you start
- 1.2 Important safety instructions and additional information
- Important information regarding safety
- Disposal of the system
- Product safety
- General care
- Accessory box
- Meter
- Power off meter
- Automatic power-off
- Shut down meter
- Automatic shutdown
- Battery Pack
- Touchscreen
- Electromagnetic compatibility (EMC)
- Electrostatic discharge (ESD)
- Wireless connectivity
- Radiofrequency radiation exposure information
- Local Area Network: protection from unauthorized access
- Wired network connection
- 1.3 System components
- 1.4 Overview of the meter
- 1.5 Overview of the code key reader
- 1.6 Overview of the Accu-Chek Inform II Base Unit
- 1.7 Overview of the Accu-Chek Inform II Base Unit Hub
- 1.8 Overview of the accessory box
- 1.9 Reagents and consumables
- 1.10 Instructions for initial setup
- 2 Powering Up and Entering an Operator ID
- 3 Patient Glucose Testing
- 3.1 Information regarding blood glucose testing
- 3.2 Performing a patient glucose test
- Overview of test procedure
- Entering or selecting the patient ID
- Entering the patient ID manually
- Selecting the patient ID from a list
- Entering a patient ID with barcode scanner
- Confirming or selecting the test strip lot
- Patient identification information
- Inserting test strips
- Obtaining a blood sample
- Applying a blood sample
- Results screen
- Adding comments
- Add barcode content to a result
- Additional Patient Test
- 4 Glucose Control Testing
- 5 Review Results
- 6 Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
- 7 Linearity Testing
- 8 Proficiency Testing
- 9 Initial Startup
- 9.1 Connecting the base unit
- 9.2 Installing or replacing the battery pack
- 9.3 Docking the meter
- 9.4 Initial setup on the meter
- 9.5 Opening the Setup Menu
- 9.6 Date and time format
- 9.7 Display options and optional tests
- 9.8 Options for test strips
- 9.9 Options for glucose control tests
- 9.10 Value ranges (normal, critical, reportable)
- 9.11 Options for Operator ID Entry
- 9.12 Patient ID options
- 9.13 Creating a setup password
- 9.14 Setting the date and time
- 9.15 Beeper options
- 10 Maintenance and Care
- 11 Troubleshooting
- 12 General Product Information
- A Appendix
- B Appendix
- C Supplement for Other Test Entry
- D Supplement for Observed Test Sequence
- E Supplement for Enhanced Workflows
- F Appendix for Accu-Chek Inform II Base Unit (legacy version)
- G Limited Warranty
- Index
144
08424705001 (04) 2021-06 USA • Accu-Chek Inform II Operator’s Manual Version 7.2
9.12 Patient ID options
This menu allows you to specify the criteria for entering a
patient ID.
1 Touch the desired option to enable it:
■ Patient Entry
– Prompt or Prompt (numeric only): The Patient ID
can be entered manually via keypad or via barcode
scanner.
– Scan only: Entry of ID via barcode scanner only.
– Prompt or List: The Patient ID can be entered
manually via keypad or via barcode scanner, or can
be picked from a list received from the DMS.
■ Patient Validation
– None: The patient ID can have any length (up to
20 characters).
– Length: The patient ID must be a specified length
(up to 20 characters). Enter the minimum and
maximum lengths after selecting this option.
– Length if numeric: Only numeric patient IDs are
validated for their length. Alphabetic patient IDs are
not validated.
– List - must be on list: The patient ID must be on a
list received from the DMS.
– List - allowed to add new: The patient ID will be
checked against a list received from the DMS.
However, it is allowed to create a new patient ID
using the meter.
2 Touch to save the settings. If the Length
option is enabled, enter the desired numbers,
or
touch to exit this menu without saving any
changes.
The display always returns to the Setup Menu (after
entering the desired numbers).
Patient ID Options
Scan only
Prompt (numeric only)
Patient Entry
Prompt
12:48 pm
03/17/20
Prompt or List
123
456
789
0
Pat. ID Min Length
8
12:48 pm
03/17/20
123
456
789
0
Pat. ID Max Length
12
12:48 pm
03/17/20
Patient ID Options
Length
None
Patient V
alidation
Length if numeric
12:48 pm
03/17/20
List - allowed to add new
List - must be on list