User's Guide

OYMotion Technologies Co., Ltd All Rights Reserved©2015-2021
9. Operation Instructions
9.1. Use of disposable electrodes: The gForceDual
TM
has 5 connecting wires: 2 pairs
(adjacent green and yellow snaps makes one pair) of EMG differential wires and a single
reference level output wire (black snap and in middle). Connect the disposable
electrodes to the wires through the snaps. Each pair of electrodes need to stick on the
skin of the interested muscle, be close to each other and along the direction of muscle
contraction. The reference level output electrode with the black snap should stick on a
bony area where muscle activity is relatively low.
9.2. EMG raw data: The gForceDual
TM
is mainly used to monitor contractions of two muscles
of the arm through the two differential pairs of EMG sensor. It is usually used to monitor
two opposite muscles, such as flexor and extensor muscles in the forearm, or bicep and
tricep muscles in the upper arm.
9.3. Powering on: When the gForceDual
TM
wrist band device is turned off, press the switch
button for 1 second to turn the device on, after, a green LED below the button flashes at
a frequency of 1Hz. There’s a short vibration as well to indicate successful powering on.
The devices will then wait for a BLE connection. Once the device establishes a BLE
connection with the Rehab Game App the green LED stays on. Once data or commands
start to transmit the green LED flashes faster at a frequency of 2Hz.
9.4. Entering firmware upgrade mode: To enter firmware upgrade mode press the button
for 15 seconds or until the green indicator light flashes 5 times rapidly first then flashes
slowly. This indicates that the gForceDual
TM
wrist band switches to firmware upgrade
mode, the BLE device name changes to gForceDual-BOOT. Entering this mode is only
recommended if the device’s firmware is damaged or the device is not working as
expected. Upon entering this mode, the user can force a firmware upgrade on the device
through the “gForce App”. When upgrade finishes, power off the device and on the next
power on cycle the device returns to normal work mode.
9.5. Powering off: When the gForceDual
TM
wrist band is turned on, press the button for 5
seconds and release to turn the device off. The green LED will flash 3 times if the internal
IMU sensor is automatically calibrated successfully during use or otherwise the device
will just turn off.
9.6. Charging: To charge the gForceDual
TM
wrist band connect the device via the USB
interface. When charging, the red LED turns on. Once charging finishes the red LED turns
off.