User Manual
Table Of Contents
- I.General Safety Requirements
- II.Safety Terms and Signs
- III.PC Configuration Requirements
- IV.Communication Interface Introduction
- V.How to Communicate Device with PC
- VI.Operation Interface of PC Software
- VII.Device Operation
- i.how to set the probe compensation
- ii. how to set the vertical system from PC softwar
- iii. how to set the horizontal system from PC soft
- iv. how to set the trigger system from PC software
- v. how to set channel from PC software
- vi. how to use automatic measurement
- vii. how to set sampling mode
- viii. how to set built-in function generator
- ix. how to use cursor measurement
- x.how to set the display system
- xi.how to change utility setting
- xii.how to use main action button
- xiii.how to use LAN communication interface
- xiv.how to work WiFi with PC software
- VIII.Technical Specifications
- IX. Appendix
56
Frequency
Sine
0.1 Hz - 5 MHz
Square
0.1 Hz - 200 kHz, with rising or falling time < 200ns
Ramp
1 Hz - 10 kHz, with rising or falling time ≥ 5μs
Pulse
1 Hz - 10 kHz, with pulse width ≥ 5μs
Amplitude
Output Amplitude
10mVpp - 5Vpp (≤5MHz) @ High Z
Amplitude Accuracy
±(1% of setting + 1 mVpp)
(typical, 1kHz sine, 0V offset)
Amplitude Resolution
10 mVpp
DC Offset Range (AC+DC)
±(2.5 Vpk - amplitude Vpp/2)
DC Offset Accuracy
±(1 % of |setting| + 1 mV + amplitude Vpp x 0.5%)
DC Offset Resolution
10 mVpp
General Part -
Communication Interface
USB Device, USB Host (hi-speed USB 2.0),
LAN (10/100Mbits)
Programming Language
SCPI
Compatibility
USBTMC, LXI, SOCKET
Power
Power Source
5V - 15V DC / 1.2A
Power Consumption
≤ 8W
Environment
Temperature
working temperature: 0 ℃ - (+40 ℃)
storage temperature: (-20 ℃) - (+60 ℃)
Relative Humidity
≤ 90%
Height
operating: 3,000 m
non-operating: 15,000 m
Cooling Method
air convection (cross-ventilation)
Mechanical
Device Dimension
w/h/d 190 x 120 x 18 mm
Weight
0.38 kg
Device Calibration Time Interval
After the device been operated for every 12 natural months (calculated from the first operation day), better
to calibrate it one time.