User Manual
Table Of Contents
- I.General Safety Requirements
- II.Safety Terms and Signs
- III.PC Configuration Requirements
- IV.Communication Interface Introduction
- V.How to Communicate Device with PC
- VI.Operation Interface of PC Software
- VII.Device Operation
- i.how to set the probe compensation
- ii. how to set the vertical system from PC softwar
- iii. how to set the horizontal system from PC soft
- iv. how to set the trigger system from PC software
- v. how to set channel from PC software
- vi. how to use automatic measurement
- vii. how to set sampling mode
- viii. how to set built-in function generator
- ix. how to use cursor measurement
- x.how to set the display system
- xi.how to change utility setting
- xii.how to use main action button
- xiii.how to use LAN communication interface
- xiv.how to work WiFi with PC software
- VIII.Technical Specifications
- IX. Appendix
34
Available in 2 display type, “Vector” and “Dots”.
Vector
to fill the room between the two adjacent sampling points, in dots of vector
Dots
to display sampling points only
The signal displayed in Vector,
In contrast with the same signal displayed in Dots,
XY Mode
Check "XY Mode", the main display area switches to 3-window VIEW mode. In the XY mode window (
upper-left window), the first channel reflects in X axis, the second channel in Y axis.
Note: When the device works in XY Mode, the record length been restricted at 1K points. The record length setting changes
to 1K points automatically.
Persistence
When “Persistence” chosen, the device emulates the persistence display effect of CRO: the originally
captured signal data in fading color, and the new data in fresh color, in cycle, continually.
The option of “Persistence” lasting available in Off, 0.5s, 1s, 2s, 5s, and Infinite, when set in “Infinite”,
the recording point continues to be there, until the persistence lasting option been changed. Press
“Clear”to clear the current persistence.