3 | iASSIST 2-Pod Version Surgical Technique
2.Preoperative Guide
Warning: The surgical procedure must take place
in an operating room theater with temperatures
ranging from 15°C to 30°C (5F to 86°F). Ensure
that the Pods are maintained in this temperature
range before being use in a surgical procedure.
Otherwise, the system will indicate that the Pods
are outside their operational temperature range
and the Pods cannot be used until they reach an
adequate temperature.
Ensure that an iASSIST Knee Pod Kit is not used
for another surgical procedure. The user must exit
and restart the application with a new pod kit to
begin another surgical procedure. A bilateral knee
procedure is considered as one procedure. As a
backup, a second iASSIST Knee Pod Kit should be
brought in the operating room theater.
The iASSIST V2 Tablet should remain in the iASSIST
V2 Tablet Shipping case during transportation.
Figure 2.0
Operationg Room Setup
Note: A full set of the implant systems
standard instrumentation is needed to
perform a surgery in conjunction with the
iASSIST system and in case the iASSIST components
cannot be used. Eects that may prevent use of the
iASSIST system include the unlikely possibility of
unresolvable inter ference eects with the wireless
communication of the system.
2.1 iASSIST V2 Tablet Setup
1. Open the iASSIST V2 Tablet Shipping Case.
2. Take out the iASSIST V2 Tablet and position it on
the Tablet Holder on a stable at surface outside
the sterile zone at a maximum distance of four
meters from the instruments and OR tables. Then
take out the top level to access the components on
the bottom level.