15 | iASSIST 2-Pod Version Surgical Technique
3.4.2 Femoral Registration
Warning: From registration to validation, the
instruments must remain stable and properly
xated to the bone to ensure accuracy of the
system.
1. Press the “Z” button on the Pod attached to the
Femoral Reference to initiate the registration
procedure.
2. Acquire 13 stable positions by accelerating and
stopping the leg to create a star-shaped pattern. An
audio feedback will be generated from the iASSIST
V2 Tablet after each acquisition until completion.
The LEDs on the Pod will blink sequentially until
completion of the acquisition.
Note: To restart the femoral registration, press
the “Z” button on the Pod attached to the
Femoral Reference or on the “Restart” button
on the tablet screen.
Figure 3.4.2
Femoral Registration
If needed, click on the “Help” button to watch a
video that explains the recommended pattern
for the femoral registration movements.
Warning: The pelvis must remain immobile
during the femoral registration.
Do not perform the femoral registration by
moving the leg in a uniaxial direction, i.e. only
in a exion/extension, abduction/adduction
and/or combination of both motions (e.g.
T-shape).
Do not perform the registration by making
circles with the leg.