Users Manual
14 | iASSIST 2-Pod Version Surgical Technique
3.4 Femur Procedure (cont)
4. Slide the Femoral Reference onto the spike.
5. Align the Femoral Reference in a neutral rotation
position.
Note: The femoral rotation may be set by:
a. Aligning the indicator arrow at the end
of the Femoral Reference to Whiteside’s line;
b. Aligning a pin inserted in the dedicated
medial-lateral hole on the body of the
Femoral Reference to the epicondyles.
6. Ax the Femoral Reference on the bone with one
3.5 x 38 mm Hex Head Screw in one of the anterior
xation holes.
7. Clip the reference pod to the femoral reference.
Note: As provided by Zimmer Biomet, the 500
RPM adaptor of the Zimmer Biomet®
Universal Power System Surgical
Instruments can be used to secure the
screw.
Figure 3.4.1
Workow Selection (cont)