Users Manual

11 | iASSIST 2-Pod Version Surgical Technique
3.2 Powering-On Pods and
Intraoperative Calibration (cont)
Warning: The calibration process including
the Pod Calibration Jig handling must be
performed wearing sterile attire.
The Pods need to be calibrated on a at,
approximately horizontal surface to ensure
optimal performance. This surface must be
sterile and free of any movement or vibrations.
4. Clip the Pods to the Pod Calibration Jig by
identifying the nose and inserting it in the
corresponding recess, then press down on the
other side of the Pod. When successfully clipped,
the status LED on the Pods will turn solid green.
5. Begin by placing face 1 of the Pod Calibration
Jig upwards on a sterile table. The numbers
corresponding to each step of the calibration are
laser-marked on the jig on the face that should
face upwards for this step. Keep the jig steady until
a sound is triggered. Continue the sequence by
positioning the jig in its remaining ve positions
(2 through 6). Wait for the conrmation sound to
conrm the acquisition between each position.
When the calibration is completed, the green
status LED of each Pod will blink.
Note: The green LEDs on the exion/extension
axis of the “Cut Guide Pod indicate the
expected position (1 through 6) of the Pod
Calibration Jig during calibration.
6. Unclip all Pods from the Pod Calibration Jig.
Figure 3.2
Powering-On Pods and intraoperative Calibration (cont)