User Guide
26 | iASSIST 2-Pod Version Surgical Technique
3.5.2 Tibial Registration
1. Ensure the “Cut Guide” Pod is clipped to the Tibial
Alignment Guide.
2. Press the “Z” button on the Pod attached to the
Tibial Alignment Guide to initiate the registration
procedure.
For steps 3-5, follow the feedback on the Pod attached
to the Tibial Alignment Guide. The iASSIST V2 Tablet
will also provide audio and visual feedback. The
number of degrees to the target range for registration
movements will be displayed on the Tablet.
Figure 3.5.2
Proximal Tibia Cut – Tibial Registration
3. Bring the leg into abduction until the green LED
stops blinking and then hold the knee steady until
the acquisition sound is heard.
4. Bring the leg into adduction until the green LED
stops blinking and then hold the knee steady until
the acquisition sound is heard.
5. Bring the leg into neutral position until the green
LED stops blinking and then hold the knee steady
until the acquisition sound is heard.
Abduction Adduction
Neutral