User Guide
34 | iASSIST 2-Pod Version Surgical Technique
Figure 3.6.1
Tibial Positioner Technique- Instrument Assembly
3. Push on the handle of the Tibial Positioner to lock
the mechanism in place.
4. Clip the “Reference” Pod to the back receptacle of
the Tibial Adjustment Mechanism.
5. Clip the “Cut Guide” Pod to the Tibial Alignment
Guide.
Workow Selection
1. Loosen the blue knob on the distal part of the
Tibial Alignment Guide.
2. Position the rod of the Tibial Alignment Guide to an
initial starting orientation per the preset position (L
or R) on the distal part of the Tibial Alignment Guide.
For a left knee procedure, the preset position is L.
For a right knee procedure, the preset position
is R.
3.6 Alternate Techniques to
Prepare Tibia Cut
The following section describes two alternate
techniques for positioning the Tibial Adjustment
Mechanism on the tibia.
3.6.1 Tibial Positioner Technique
Instrument assembly
Note: The Tibial Positioners are side specic
and must be used with the corresponding
Tibial Adjustment Mechanism. Ensure to pick
the appropriate instruments according to the
patient side.
1. Pull out the handle of the Tibial Positioner.
2. Connect the Tibial Positioner to the Tibial
Adjustment Mechanism while still pulling on the
handle.