Cleaning Instructions

53 | iASSIST 2-Pod Version Surgical Technique
5.2 iASSIST Knee Master Instruments Kit
Additional specic cleaning instructions as dened in the text are indicated as applicable to each
instrument as follows:
A
Requires disassembly
B
Requires water jet to ush dicult to access areas
C
Screw/unscrew components while ushing the area
D
Requires a minimum 10 minutes ultrasonic cleaning cycle in an enzymatic solution
E
Screw/mechanism should be checked and lubricated
with a medical grade surgical lubricant normally
after each cleaning as determined upon inspection
Description Product ID Qty
Sterilization and Specic
Cleaning Instructions Additional Notes
Femoral Spike
20-8011-051-00 1 Autoclave
Tibial Alignment Guide
20-8011-013-00 1 Autoclave
Additional specic cleaning
requirements:
A
B
C
D
E
A
B
C
B