Table 5.9: Guidance and Manufacturers Declaration – Electromagnetic Immunity
The pods are intended for use in the electromagnetic environment specied below.
The customer or the user of the pods should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2 test level
Compliance level
Conducted RF
IEC 61000-4-6
3 VRMS
0,15 MHz – 80 MHz
6 VRMS in ISM bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
Not applicable
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
3 V/m
80 MHz – 6 GHz
80 % AM at 1 kHz
Proximity elds from RF wireless
communications equipment IEC
61000-4-3
See table 5.10 See table 5.10
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any part of the iASSIST Pod, including cables specied by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Table 5.10: Guidance and Manufacturers Declaration – Electromagnetic Immunity
The pods are intended for use in the electromagnetic environment specied below.
The customer or the user of the pods should assure that it is used in such an environment.
Immunity test Test Frequency Modulation
IEC 60601-1-2 test level
Compliance level
Proximity eld from RF
wireless communication
equipment IEC 61000-4-3
385 MHz Pulse Modulation : 18 Hz 27 V/m 27 V/m
450 MHz
FM ± 5 Hz deviation: 1kHz sine
X
Pulse Modulation : 18Hz
28 V/m 28 V/m
710 MHz
745 MHz
780 MHz
Pulse Modulation : 217Hz 9 V/m 9 V/m
810 MHz
870 MHz
930 MHz
Pulse Modulation : 18 Hz 28 V/m 28 V/m
1720 MHz
1845 MHz
1970 MHz
Pulse Modulation : 217 Hz 28 V/m 28 V/m
2450 MHz Pulse Modulation : 217 Hz 28 V/m 28 V/m
5240 MHz
5500 MHz
5785 MHz
Pulse Modulation : 217 Hz 9 V/m 9 V/m