Table 5.6: Guidance and Manufacturers Declaration – Electromagnetic Immunity
The iASSIST V2 Tablet is intended for use in the electromagnetic environment specied below.
The customer or the user of the iASSIST V2 Tablet should assure that it is used in such an environment.
Immunity test Test Frequency Modulation
IEC 60601-1-2 test level
Compliance level
Proximity eld from RF
wireless communication
equipment IEC 61000-4-3
385 MHz Pulse Modulation : 18 Hz 27 V/m 27 V/m
450 MHz
FM ± 5 Hz deviation: 1kHz sine
X
Pulse Modulation : 18Hz
28 V/m 28 V/m
710 MHz
745 MHz
780 MHz
Pulse Modulation : 217Hz 9 V/m 9 V/m
810 MHz
870 MHz
930 MHz
Pulse Modulation : 18 Hz 28 V/m 28 V/m
1720 MHz
1845 MHz
1970 MHz
Pulse Modulation : 217 Hz 28 V/m 28 V/m
2450 MHz Pulse Modulation : 217 Hz 28 V/m 28 V/m
5240 MHz
5500 MHz
5785 MHz
Pulse Modulation : 217 Hz 9 V/m 9 V/m