Table 5.6: Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The iASSIST V2 Tablet is intended for use in the electromagnetic environment specied below.
The customer or the user of the iASSIST V2 Tablet should assure that it is used in such an environment.
Immunity test Test Frequency Modulation
IEC 60601-1-2 test level
Compliance level
Proximity eld from RF
wireless communication
equipment IEC 61000-4-3
385 MHz Pulse Modulation : 18 Hz 27 V/m 27 V/m
450 MHz
FM ± 5 Hz deviation: 1kHz sine
X
Pulse Modulation : 18Hz
28 V/m 28 V/m
710 MHz
745 MHz
780 MHz
Pulse Modulation : 217Hz 9 V/m 9 V/m
810 MHz
870 MHz
930 MHz
Pulse Modulation : 18 Hz 28 V/m 28 V/m
1720 MHz
1845 MHz
1970 MHz
Pulse Modulation : 217 Hz 28 V/m 28 V/m
2450 MHz Pulse Modulation : 217 Hz 28 V/m 28 V/m
5240 MHz
5500 MHz
5785 MHz
Pulse Modulation : 217 Hz 9 V/m 9 V/m