Table 5.5: Guidance and Manufacturers Declaration - ELECTROMAGNETIC IMMUNITY
The iASSIST V2 Tablet is intended for use in the electromagnetic environment specied below.
The customer or the user of the iASSIST Tablet should assure that it is used in such an environment
IMMUNITY test
IEC 60601-1-2 test
Level
Compliance
level
Conducted RF
IEC 61000-4-6
3 Vrms
0.15 MHz - 80 MHz
6 VRMS in ISM
bands between
0.15 mHz and 80
Mhz
80% AM at 1 kH
3 Vrms
0.15 MHz - 80
MHz
6 VRMS in ISM
bands between
0.15 mHz and
80 Mhz
80% AM at 1 kH
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
3 V/m
80 MHz - 2,7
GHz
80% AM at 1
kHz
Proximity elds
from RF wireless
communications
equipment IEC
61000-4-3
See Table 5.6 See Table 5.6
Warning:
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be
user no closer than 30 cm (12 inches) to any part of the iASSIST Tablet, including cables specied by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Electrical fast transient/burst-
IEC 61000-4-4
± 2 kV for input power port
(a.c. and d.c)
100 kHz repetition frequency
± 1 kV for signal input/output
parts
100 kHz repetition frequency
± 2 kV for input power port
(a.c. and d.c)
100 kHz repetition frequency
± 1 kV for signal input/output
parts
100 kHz repetition frequency
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5
± 0.5 kV, ± 1 kV line-to line for
input power port (a.c. and
d.c.)
± 0.5 kV , ± 1 kV , ± 2 kV line-to-
ground for input power port
(a.c. and d.c.)
± 2 kV line-to-ground for signal
input/output parts
± 0.5 kV, ± 1 kV line-to line for
input power port (a.c. and d.c.)
± 0.5 kV , ± 1 kV , ± 2 kV line-to-
ground for input power port
(a.c. and d.c.)
± 2 kV line-to-ground for signal
input/output parts
Mains power quality should be
that of a typical commercial or
hospital environment.